Calculation of iris-claw IOL power for correction of late in-the-bag IOL complex dislocation

Visualitza/ Obre
Data de publicació
2017Citació recomanada
Huerva, Valentín;
Ascaso, Francisco J.;
Caral, Isabel;
Grzybowski, Andrzej;
.
(2017)
.
Calculation of iris-claw IOL power for correction of late in-the-bag IOL complex dislocation.
BMC Ophthalmology, 2017, vol. 17, p. 122.
https://doi.org/10.1186/s12886-017-0516-1.
Metadades
Mostra el registre d'unitat completResum
Background: To assess the constants and formula for aphakia correction with iris-claw IOLs to achieve the best refractive status in cases of late in-the-bag IOL complex dislocation. Methods: A literature search was performed. The following data were obtained: Iris-claw IOL model, Iridal or retroiridal enclavation, A-constant, ultrasound or optical biometry, formula employed and refractive outcomes. Acceptable emmetropia was considered if the resulting spherical equivalent (SE) was within ±1.00 D. Results: The majority of the studies used SRK/T formula (66.6%). The 88.9% of the reports obtained a SE within ±1.
00 D. Using A-115 for ultrasound biometry and A-115.7 for optical biometry and SRK/T formula, the emmetropia (±1.00 D) of SE, was able to get near 100% of reported cases over the pupil implantation. However, the emmetropia decreased to 80% when the enclavation is retropupilar using the same formula. The A-constant can vary from 116.7 to 117.5 for retropupilar enclavation. Conclusions: Using A-115 for ultrasound biometry and A-115.7 for optical biometry and SRK/T formula, ±1.00 D of SE, is able to get near 100% of cases. Nevertheless, ±1.00 D of SE decreased to 80% of the cases when the enclavation is retropupilar.
És part de
BMC Ophthalmology, 2017, vol. 17, p. 122Projectes de recerca europeus
Col·leccions
Els fitxers de llicència següents estan associats amb aquest element: